Introduction:
This cyclone is intended to be used with any sampling pump meeting the following pressure/volume
characteristics:
Media Type37mm
Flow RateLpm
PressureDrop
Cm of H2o
GF/A Fiberglass 2.2 17.13
QMA Quartz 2.2 13.14
Pallflex Tissue Quartz 2.2 17.10
Pall Teflo 2 µm 2.2 18.42
0.8 µm MCE 2.2 18.80
Operation:
Remove the front portion of a 3-piece cassette (blue plug) by inserting a coin or opener in the
annular slot closest to the blue plug and twisting. Save the portion removed to reinstall, for filter
protection after sampling. Press the portion of the cassette containing the filter onto the tapered
top of the cyclone – fitted with an orange “O” ring.
Next, remove the red plug from the back of the cassette and push in the taper Luer plug, which is
attached to the hose.
The cyclone is clipped to the subjects lapel and the long hose is run over the subject’s shoulder to
the personal sampler appropriately situated on the waist belt.
Proceed with sampling according to your experience, regulations, or the instructions for the personal
sampler being used. The procedures are the same as for a routine cassette dust sample, except
that only the fraction based on the flow rate reaches the filter surface. Selection of filter materials
varies with different countries, regulations or research intentions.