Two-component, 100%-solids epoxy system for
generalpurpose stress analysis. Transparent, medium
viscosity. Cure time as low as six hours at +125°F (+50°C).
Recommended for more critical applications, including
transducers. It has a longer pot life than M-Bond AE-10
which allows more time for multiple gage installations.
Strain Gage Adhesive
Elevated-temperature postcure is recommended for
maximum stability, and/or tests above room temperature.
Highly resistant to moisture and most chemicals,
particularly when postcured. For maximum elongation,
bonding surface must be roughened. Cryogenic
applications require very thin gluelines.
CHARACTERISTICS
PARAMETER DETAILS
OPERATING TEMPERATURE
RANGE
Long Term: –452° to +200°F (–269° to +95°C).
Transducers: to +175°F (+80°C).
ELONGATION CAPABILITIES 2% at –320° (–195°C), 10% to 15% at +75°F (+24°C), 15% at +200°F (+95°C).
SHELF LIFE
Minimum 12 months at +75°F (+24°C); or 18 months at +40°F (+5°C). If crystals form in resin jar,
heat to +120°F (+50°C) for 30 minutes. Cool before mixing.
Refer to product label for most recent information.
POT LIFE 1-1/2 hours at +75°F (+24°C).
CLAMPING PRESSURE 5 to 20 psi (35 to 140 kN/m2
).
CURE REQUIREMENTS*
Recommended Postcure:
2 hours at 25°F (15°C) above
maximum operating
temperature.
Optimum Performance
Transducer Postcure:
1 hour at +200°F (+95°C).