Lighthouse 1100粒子传感器简单通信介绍和各引脚功能介绍

Check data. For the zero count test procedure, please refer to the Operating Manual.
下载次数:1 浏览次数:880发布时间:2018-10-27 20:03
允许查看的会员等级:普通会员   高级会员   VIP会员   

文档详情

1.Unpack instrument and inspect for shipping damage.

2.Verify that site cable connector pin outs are correct.

3.Verify AC power is correct - do NOT apply power, yet.

4.Remove the Inlet and Outlet protective caps and attach the

Purge filter to REMOTE Inlet nozzle.

5.Verify that vacuum source meets minimum requirements,

attach vacuum tubing to REMOTE and start vacuum flow.

When power is applied in Step 6, flow LED should turn ON

6.Apply power to instrument and verify that Power LED turns

ON.  Run the unit for fifteen minutes to warm up the LASER.

7. Check the Service LED - if it remains ON, proceed to Step 9.

8.Check data.  For the zero count test procedure, please refer to the Operating Manual.

9.For any failure, power OFF and contact Lighthouse Support.