2.4.2 Starting a new measurement
Leave the stand-by mode and start a new measurement by pressing the [Stand-by] key.
Setting the measurement mode and choice of the displayed size channels
Holding the key {Size Mode} and pressing additionally the key {+} or {-} lets you
change the measurement mode, either particle mass or particle concentration. The
LCD display shows on the right side the accordant unit µg/m³ or particle per liter as
"/l".
Decisive for the measurement mode and the change of the measurement mode is
the setting in the upper line of the LCD display. Leaving the size range by pressing
the key [+] or [-] the measurement mode changes.
Example 1: The display shows in the mass mode >20µm …..µg/m³ in the upper line. Pressing {+}
again and the measurement mode will change to particle concentration, indicated by >0.3µm …..
/l.
Example 2: The display shows >0.25µm….. /l in the concentration mode, pressing {-} will switch to
mass mode, indicated by >32µm….. µg/m³. The size channel that is chosen in the lower line of the
LCD has no influence on the choice of the measurement mode. It simply shows the data of a second
channel within operational mode.
Please notice that the data sent via the RS-232 interface and the data stored onto the
storage card are the same type as the set displayed operation type. That means either dust
mass or particle concentration can be displayed and stored!
Displaying the analog values
Pressing only the key {Size/Mode} longer than three seconds or pressing
additionally the key {Date/Time}, the values of the analog input values and time will
be displayed. This remains set until pressing any other key.
Requesting the battery capacity and measurement location number
By pressing the {Battery/Location} key the current battery capacity as percentage
as well as the set measurement location number will be displayed. When operating
with a mains adapter, always 130% are displayed. At a fresh charged battery values
above 100% are possible. Pressing and holding the key {Battery/Location} the
measurement location number can be altered by means of the keys {+} or {-}. The
measurement location number is an additional differentiation of various
measurements and can be set between 1 and 99.
Model 11-A
GRIMM AEROSOL TECHNIK GmbH & Co.KG Page 23 of 80
When reading out the data storage card via the Windows software, the datasets can be
chosen individually according to their measurement location number, which shortens and
simplifies the data transmission.
Changing the alarm thresholds, necessity of a diluter at coincidence
Simultaneous pressing of the keys {Alarm} and {+} or {-} changes the alarm
thresholds. A value of 0 deactivates the alarm function. A value bigger than 0
activates the alarm. The alarm value always refers to the size channel in the upper
line of the LCD and the set operational mode (mass or particle concentration).